COMPONENTS AND CONTROLS (cont.)
Leveled
Not leveled
Shipping thumb
screw(located on
bottom of scale)
REMOVE ALL SHIPPING SCREWS
BEFORE USING SCALE, TWO
SCREWS FROM TOP AND THUMB
SCREW FROM BASE.
8529-B222-08 REV.A
Turn the top two (2)
shipping screws
counter-clockwise to
remove.
2. PLACEMENT
Place the scale on a stable,
vibration-free location away from
direct sunlight and away from any
rapid moving air source. Make
certain the power cord is stored
out of the way of normal traffic.
CAUTION!!! DO NOT
place the scale on
any unstable cart,
stand or table. The
scale may fall, causing injury to
the operator, and seriously
damage the unit; or proper
operation of the scale may be
inhibited.
3. LEVEL SCALE
Before mounting the commodity tray onto the scale, level the scale if necessary by adjusting the
four legs on scale bottom, up or down, until the bubble in the level gage located on the scale
housing is centered as shown in the figure above. After level is achieved, lock legs in place by
tightening lock nuts on legs against the scale bottom.
POWER SUPPLY
1. Plug power supply into a wall receptacle that supplies 115 VAC power and into the connector on
the underside of the chassis.
2. On models requiring 230 VAC power, use the Cardinal/Detecto model 8529-B216-08, 230 VAC
power supply. After installation of the proper connection, plug scale power supply into 230 VAC
receptacle.
CAUTION!!! - To avoid electrical hazard and severe damage to the scale, use only
the wall-plug-in power module provided with the scale.
3. START-UP - The scale is equipped with ON and OFF keys on the keypad. When the ON key is
pushed, the scale will show zero weight, signalling that the scale is ready for use. If the deadload
on the scale is not within ± 4% of the deadload used during calibration, the scale will display LoAd
or UnLoAd, indicating that weight must be added or removed, respectively, from the platform. Once
the deadload is within ± 4% of the deadload used during calibration, the scale will zero and is ready
for use.